Determination of the content of migratory elements in toys - Database & Sql Blog Articles

WIFI test head can be customized for other specifications
Test - lowercase jpg
Kaixin micro test

天瑞ICP2060T

Inductively coupled plasma atomic emission spectrometer

Measurable elements:

Aluminum, barium, arsenic, antimony, boron, cadmium, cobalt, copper, lead, manganese, mercury, nickel, selenium, antimony, tin, zinc, etc.

70

A variety of elements.

Reference standard

:

EN71-3 (

In this standard, only the sample processing and data processing are explained and specified. No instrument or reagent is specified, but the detection limit of the instrument should be less than one tenth of the limit standard.

)

principle:

The simulated material is continuously in contact with gastric acid for a period of time after swallowing, thereby determining the concentration of soluble elements. (simulation conditions:

37±2

°C

, 0.07mol/L

hydrochloric acid)

Instrument introduction:

ICP2060T

Inductively coupled plasma emission spectrometer is used to determine the content of constant, trace and trace elements in various substances (soluble in hydrochloric acid, nitric acid, hydrofluoric acid, etc.). The instrument has the characteristics of high efficiency, strong anti-interference, high degree of automation, easy operation, stable and reliable, wide test range, fast analysis speed and low detection limit.

ICP2060T

Measurable

70

Multiple elements, measurable in one minute

5-8

Elements, and can achieve multi-element simultaneous analysis, customers can freely choose the number of elements and arrange the measurement order. Widely used in the detection of transportable heavy metals in toys.